Recall Z-0823-05 ยท NBW System, Test, Blood Glucose, Over The Counter

Lifescan Inc

LifeScan Brand InDuo Blood Glucose Meter

DateMay 17, 2005 (initiated Apr 11, 2005)
StatusTerminated
ReasonUser may inadvertently change the Units of Measure from mg/dL to mmol/L and the blood glucose results could be misinterpreted. This may lead to under treatment and potential for hyperglycemia.
Root causeOther
ActionOn 4/11/05, the firm issued letters via Federal Express to all its direct consignees, informing them of the affected products and providing instructions on the correction. Owner's booklet: Firm contacted all accounts by letters of 11/21/03, 2/4/2004, 2/17/04 and 2/9-10/2004 providing important information regarding the owner's booklet.
DistributionDistribution of the product is worldwide. ***Distribution in the United States was between January 15, 2001 for OneTouch Ultra Blood Glucose Meter; April 13, 1998 for OneTouch FastTake Blood Glucose Meter and November 30, 2001 for InDuo Blood Glucose Meter.*** The total number of quantities shippe
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K024194ONE TOUCH ULTRA AND ONE TOUCH INDUO BLOOD GLUCOSE MONITORING SYSTEMS