ClearanceBrief
Recall Z-0883-05 ยท CGA Glucose Oxidase, Glucose

Roche Diagnostics Corp.

Roche brand AccuData GTS/GTS Plus System Replacement Base/Module; Catalog number 3426050, material number 03426050001.

DateJun 11, 2005 (initiated Apr 18, 2005)
StatusTerminated
ReasonIf foil wrapped packages or other metallic objects are lodged behind the supply drawer, an electrical short may result and high heat generated; this could result in plastic melting and battery leakage for customers using 'C' cell batteries to power their base and could render the system inoperable (system won't turn on) for customers using the A/C adaptor without batteries.
Root causeOther
ActionConsignees were notified via letter dated 4/18/05, which included instructions on how to inspect their unit for damage, and how to prevent future damage.
DistributionNationwide and Guam.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K924475ACCUDATA GLUCOSE TEST STATION
Boehringer Mannheim Corp., Oct 13, 1992