Recall Z-0841-05 ยท FRN Pump, Infusion

Medrad Inc

Continuum MR Compatible Infusion Pump. Pump Model #3009135

DateMay 13, 2005 (initiated Mar 22, 2005)
StatusTerminated
ReasonCustomer complaint that a continuum infusion pump was attracted to an MR scanner due to customer not completely engaging the pump into the bracket.
Root causeOther
ActionThe recalling firm issued a Field Correction letter dated 3/9/05 to all their customers. The letter instructed the consignee on how to apply the new warning labels, to discard the old Operation Manual and replace it with the new one.
Quantity461 units
DistributionThe products were shipped nationwide to hospitals and outpatient MRI office. The product was shipped to government facilities in CA, DC, MD, NJ, SC, WA, and WV. The product was also shipped to Austriala, Canada, Mexico, United Arib, Austria, Germany, Denmark, France, Great Britain, Hong Kong, Icela
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021988MEDRAD CONTINUUM MR COMPATIBLE INFUSION SYSTEM
Medrad, Inc., Aug 29, 2002
K032771MEDRAD CONTINUUM MR INFUSION SYSTEM
Medrad, Inc., Sep 23, 2003