Recall Z-0788-05 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Asahi Medical Co Ltd

Asahi AM-BIO Series Hollow Fiber Dialyzers; a wet model multiple use dialyzer consisting of hollow fiber membranes of alkyl ether polymer grafted cellulose housed within a plastic casing of styrene butadiene block polymer, filled at the factory with fluid to facilitate priming by the user, and sterilized by gamma radiation before shipment; Asahi Medical Co., Ltd., 9-1, Kanda Mitoshirocho, Chiyoda-

DateMay 3, 2005 (initiated Mar 31, 2005)
StatusTerminated
ReasonExposure to freezing temperatures may damage the hollow fiber dialyzers and result in blood leaks during dialysis.
Root causeOther
ActionAsahi sent Henry Schein, Inc. a recall letter dated 3/30/05, requesting the recall of the affected lots of Asahi labeled product, and Henry Schein, Inc. recalled the Asahi labeled dialyzers by letter dated 3/31/05, with a copy of the Asahi 3/30/05 letter. The accounts were informed that the dialyzers may have been exposed to temperature variances during distribution, and werw requested to cease distribution/use of the specific lots listed and return them to Henry Schein for credit. The recall was expanded to include the Terumo labeled dialyzers distributed by Asahi by Henry Schein letter dat
Quantity95,129 units
DistributionNationwide and Puerto Rico
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983720ASAHI AM-BIO SERIES DIALYZERS (WET MODELS)& (DRY MODELS), MODELS AM-BIO-50,AM-BIO-65, AM-BIO-75, AND AM-BIO-100, AM-BIO-
Asahi Medical Co., Ltd., May 17, 1999