Recall Z-0709-05 ยท NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)

St Jude Medical

Implantable Cardioverter-Defibrillator Models affected include the following: Atlas+ Models V-193, V-193C and V-243 Epic+ Models V-196, V-236 and V-239 Epic Models V-197 and V-233 Atlas Model V-242 Epic HF Models V-337 and V-338 Epic+HF Models V-339 and V-350 Atlas+ HF Models V-340, V-341 and V-343

DateApr 13, 2005 (initiated Mar 10, 2005)
StatusTerminated
ReasonSensitivity of a magnetic switch was not being set correctly. This could result in delivery of an unneeded shock during surgery with use of electrocautery.
Root causeOther
ActionNotification materials were distributed to sales representatives on March 10, 2005. They will visit each doctor that is following the devices.
Quantity1488
DistributionNationwide, Canada, Australia, Sweden, Brazil, Colombia, Argentina.