Recall Z-1392-05 ยท JHI Visual, Pregnancy Hcg, Prescription Use

Abbott Laboratories MPG

ABBOTT TestPack +Plus hCG Combo with On Board Controls (OBC), list number 7B34-16; the kit contains 20 reaction discs, Anti-beta hCG (Goat) and Anti-alpha hCG (Mouse, Monoclonal) coated; Abbott Laboratories, Abbott Park, IL 60064 USA

DateAug 24, 2005 (initiated Mar 16, 2005)
StatusTerminated
ReasonFalse Positive results may be obtained with the identified lots of list 7B34-16 when using a negative patient sample or a negative control.
Root causeOther
ActionAbbott notified their affiliates in all four countries via e-mail on 3/16/05. Each Abbott international country organization is responsible for translating the customer letter, contacting customers and determining the method of effectiveness checks based on their country requirements. The recall letters inform the accounts of the false positve results, and were requested to destroy any remaining inventory of the affected lots.
Quantity7607 kits
DistributionCanada, Germany, Hong Kong and Australia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K965116ABBOTT TESTPACK PLUS HCG COMBO WITH OBC
Abbott Laboratories, Feb 20, 1997