Recall Z-0816-05 ยท IYB Tube Mount, X-Ray, Diagnostic

Del Medical Systems Group

Tube Stand for the Dynarad Phantom Mobile X-Ray System; Del Medical Systems, 11550 West King Street, Franklin Park, IL 60131

DateMay 11, 2005 (initiated Mar 10, 2005)
StatusTerminated
ReasonThe Phantom Portable X-Ray System Stand may have a suspect weld supporting the bracket holding the locking cylinder assembly. It the weld fails, the tubehead arm assembly could fall and contact the patient or operator.
Root causeOther
ActionDel Medical sent Field Correction Recall letters dated 3/10/05 to the direct accounts, informing them of the potential for a defective weld supporting the braket holding the locking cylinder assembly. If the weld fails, the tubehead arm assembly could potentially fall and contact the patient or operator. The accounts were requested to locate the affected systems and take them out of service until they are inspected and, if needed, repaired.
Quantity31 stands
DistributionNationwide, Mexico and Thailand.