Recall Z-1046-05 ยท MKJ Automated External Defibrillators (Non-Wearable)

Heartsine Technologies Inc

HeartSine Samaritan AED Defibrillators Models: SAM 001, SAM 002, SAM 003.

DateJul 28, 2005 (initiated Oct 5, 2005)
StatusTerminated
ReasonImmediately after battery installation the devices announce a low battery fault warning and shut themselves off, or the service indicator icon starts blinking indicating a fault condition.
Root causeDevice Design
ActionAED software revision 1.0.20 and higher incorporate software enhancements to make the AED more forgiving for extreme variations in user technique. Firm will provide distributors with field correction notice dated 10/5/05. Customers to be notified by mid-October 2005.
Quantity1929 units
DistributionNationwide, Puerto Rico, Argentina, Australia, Austria, Belgium, Denmark, Ecuador, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Netherlands, Pakistan, Phillipines, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Thaila
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023854SAMARITAN AED, MODEL SAM-01
Heartsine Technologies, Inc., Feb 14, 2003