ClearanceBrief
Recall Z-0664-05 ยท DQA Oximeter

Nellcor Puritan Bennett

Nellcor brand OxiMax Pulse Oximeter, model number N-595 Made in Ireland; Distributed by Tyco Healthcare Group, LP, Nellcor Puritan Bennett, Inc., Pleasanton, CA 94588 U.S.A.

DateMar 30, 2005 (initiated Oct 18, 2004)
StatusTerminated
ReasonThe internal speaker components are failing and do not produce an audio alarm in the event of patient emergency.
Root causeOther
ActionCustomers of record were notified of the advisory by letter sent via US Mail on October 18, 2004, via certified mail. A second letter was sent on December 9, 2004, advising customers that upgrades would be available. Upgrades would be available via returns to the firm''s service center, upgrade kits and speaker replacements via self-installation, and a third party service provider.
Quantity42,451 units
DistributionThere are 2,950 Direct US and US territory Customers. Product shave also been distributed to Canada, Europe, Japan, Latin America and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012891OXIMAX PULSE OXIMETRY SYSTEM WITH N-595 PULSE OXIMETER AND OXIMAX SENSORS AND CABLES (AKA ACCESSORIES)
Nellcor Puritan Bennett, Inc., Mar 7, 2002