Recall Z-0662-05 ยท DQA Oximeter

Nellcor Puritan Bennett

Nellcor brand Pulse Oximeter, Model NPB290, MALLINCKRODT, Made in Ireland, Oxismart, Made in Ireland; Distributed by Nellcor Puritan Bennett, Inc. Pleasanton, CA 94588 U.S.A.

DateMar 30, 2005 (initiated Oct 18, 2004)
StatusTerminated
ReasonThe internal speaker components are failing and do not produce an audio alarm in the event of patient emergency.
Root causeOther
ActionCustomers of record were notified of the advisory by letter sent via US Mail on October 18, 2004, via certified mail. A second letter was sent on December 9, 2004, advising customers that upgrades would be available. Upgrades would be available via returns to the firm''s service center, upgrade kits and speaker replacements via self-installation, and a third party service provider.
Quantity7,814 units
DistributionThere are 2,950 Direct US and US territory Customers. Product shave also been distributed to Canada, Europe, Japan, Latin America and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K972837NPB-290 PULSE OXIMETER (NPB-290)
Nellcor Puritan Bennett, Inc., Oct 21, 1997