Recall Z-0864-05 ยท LXH Orthopedic Manual Surgical Instrument

Zimmer Inc.

Zimmer brand fracture blade plate instrument Chisel, 95 degree, small; assembly part 4705-03-10, catalog number 00-4705-003-10.

DateJun 2, 2005 (initiated Feb 9, 2005)
StatusTerminated
ReasonThe engraved calibration lines are incorrectly specified, which will result in the user obtaining a reading that the chiseling removed material to a depth 10mm deeper than what was actually removed, thus guiding the user to select a blade plate that is longer than will fit.
Root causeOther
ActionThe firm's distribution network was notified via email on 2/9/05, instructed to check their inventory and that of their customer hospitals, and to return the product to Zimmer.
DistributionNationwide and Canada.
FDA recordOpen on fda.gov