Recall Z-0638-05 ยท MRG System, Test, Vitamin D

Diasorin Inc.

Liaison 25 OH Vitamin D Assay, Manufacturer part #310900, Lot #114668

DateMar 18, 2005 (initiated Feb 14, 2005)
StatusTerminated
ReasonAn internal inspection indicates that Calibrator 2 and Conjugate are switched in their positions in the integral. This will result in Calibrator 1 RLU's similar to that of background RLU, and Calibrator 2 RLU's near expected range. A curve can not be calculated from this data.
Root causeOther
ActionDiaSorin Customer Notification dated 02/14/05 was sent to customers instructing them to destroy all kits from the affected lot number. Customers were also asked to complete and return a form indicating how many kits were destroyed.
Quantity31 kits
DistributionMA, Spain, Germany & Belgium
FDA recordOpen on fda.gov