Recall Z-0565-05 ยท KCY Tourniquet, Pneumatic

Zimmer Orthopaedic Surgical Products

Zimmer Orthopaedic Surgical Products, Disposable Cuff, Single Port, Double Bladder Cuff, Sterile, Zimmer, Dover, OH USA. The cuff is a sterile, single use product equipped with an integral fill line. The product is packed 10 cuffs per unit.

DateFeb 25, 2005 (initiated Feb 11, 2005)
StatusTerminated
ReasonThe port/tube bond in the affected cuffs may separate during surgery causing the pressure in the cuff to release.
Root causeOther
ActionConsignees were notified by letter on 02/11/2005.
QuantityTotal 430 cuffs
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.