Recall Z-0651-05 ยท LCP Assay, Glycosylated Hemoglobin

Roche Diagnostics Corp.

Roche ULTIMATE 5 HBA1C test kit for use with MIRA analyzers; catalog number 20755656322.

DateMar 23, 2005 (initiated Feb 15, 2005)
StatusTerminated
ReasonThe product is not uniform across the cassettes of some lots, which will result in an inconsistent bias of patient results.
Root causeOther
ActionRecall letters dated 2/15/05 were sent to each user instructing customers to run controls before the use of each cassette or kit in a lot, instead of once per lot as previously instructed, and to notify the physician at the facility.
Quantity1800 kits.
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K952337ROCHE UNIMATE 3/UNIMATE 5 HBA1C, HBA1C CALIBRATOR, HBA1C CONTROLS N & P
Roche Diagnostic Systems, Inc., Feb 9, 1996