Recall Z-0653-05 ยท GWQ Full-Montage Standard Electroencephalograph
Compumedics Usa, Ltd
Nonin Xpod Oximeter model numbers 3011 and 3012, distributed for use with the Compumedics E-Series Digital Amplifier.
| Date | Mar 24, 2005 (initiated Feb 24, 2005) |
| Status | Terminated |
| Reason | Under specific conditions, the housing of the pulse oximeter contected to the amplifier can become excessively warm causing a burn upon contact with skin. |
| Root cause | Other |
| Action | The firm initiated the recall on February 24, 2005 via letter with return response post card. |
| Quantity | 1376 units |
| Distribution | Product was distributed nationwide and to the following foreign countries: Australia, Canada, Greece, Hong Kong, Hungary, India, Iran, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Pillippines, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Taiwan, Thailand, Turkey, |