Recall Z-0652-05 ยท GWQ Full-Montage Standard Electroencephalograph

Compumedics Usa, Ltd

Compumedics E-Series Digital Amplifier System part number 8008-0001-01 and 8008-0001-02, manufactured by Compumedics Limited, Australia

DateMar 24, 2005 (initiated Feb 24, 2005)
StatusTerminated
ReasonUnder specific conditions, the housing of the pulse oximeter contected to the amplifier can become excessively warm causing a burn upon contact with skin.
Root causeOther
ActionThe firm initiated the recall on February 24, 2005 via letter with return response post card.
Quantity1376 units
DistributionProduct was distributed nationwide and to the following foreign countries: Australia, Canada, Greece, Hong Kong, Hungary, India, Iran, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Pillippines, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Taiwan, Thailand, Turkey,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K000068COMPUMEDICS E-SERIES EEG SYSTEM
Compumedics Neuro Science Pty , Ltd., Mar 13, 2000