Recall Z-0637-05 ยท MRG System, Test, Vitamin D

Diasorin Inc.

LIAISON 25 OH Vitamin D Assay, Manufacturer part # 310900, Lot # 114439E.

DateMar 18, 2005 (initiated Feb 14, 2005)
StatusTerminated
ReasonResults of an internal inspection show Lot 114439E to have calibrator 1 and Calibrator 2 switched in their position in the integral. This results in an inability to calculate a curve, due to the reversal of calibrator values.
Root causeOther
ActionDiaSorin Customer Notification dated 02/14/05 was sent to customers instructing them to destroy all kits from the affected lot number. Customers were also asked to complete and return a form indicating how many kits were destroyed.
Quantity131
DistributionMI, NJ, Germany, Spain and Belgium
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032844LIAISON 25 OH VITAMIN D
DiaSorin, Inc., Feb 12, 2004