ClearanceBrief
Recall Z-0607-05 ยท

Roche Diagnostics Corp.

Roche brand ISE SnapPak for use only in the AVL 9110/9140 Electrolyte Analyzer Assembled and distributed by Roche Diagnostics Corp., Indpls., IN.; Ref/No. 03112314018, Part number BP5032.

DateMar 11, 2005 (initiated Dec 21, 2004)
StatusTerminated
ReasonA defective seam in the two chamber pouch may result in the mixing of reagents, which would either result in an inability to calibrate the analyzer or would bias patient results.
Root causeOther
ActionOwn name distributor consignees were notified on or about 1/3/05 via letter dated 12/21/04, and were requested to discontinue shipment of the product, notify all of their customers of the recall and to discard product. Roche consignees were notified via letter dated 12/21/04, requested to notify their customers if they had further distributed the product, and instructed to discard all inventory of product.
Quantity9280 units
DistributionNationwide, Argentina, Austria, Brazil, Chile, China, Colombia, Ecuador, France, Germany, Guatemala, India, Indonesia, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, New Zealand, Panama, Philippines, Poland, Russia, San Salvador, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thai
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K932642AVL 9110PH ANALYZER
Avl Scientific Corp., Oct 14, 1993
K972673AVL 9181 ELECTROLYTE ANALYZER
Avl Scientific Corp., Aug 20, 1997