Recall Z-0807-05 ยท DHN Antinuclear Antibody, Indirect Immunofluorescent, Antigen, Control

Diasorin Inc.

Programming software (all versions) for Polymedco PolyTiter Immunofluorescent Titration System and used with DiaSorin ANAFLUOR Indirect Fluorescent Antibody Test.

DateMay 10, 2005 (initiated Sep 24, 2004)
StatusTerminated
ReasonA problem with software for the PolyTiter Immunofluorescent Titration System in that no flags or warnings appear when PolyTiter Calibrator Values are re-entered or changed (including values changed to unacceptable values) after the initial curve has been generated, if the 'refresh button' has not been pressed. The graph may appear valid; however erroneous results may occur.
Root causeOther
ActionCustomers were sent a Technical Bulletin describing the correct procedure to be used when entering data into the Poly/Titer tiration system, until the updated software is installed
Quantity13
DistributionOH, NJ, KY, TX, NC, FL, VA, WA, AL, CO, MO, AZ , CA , CT & NV
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K822397ANA HEP-2 FLUORO-KIT
Clinical Sciences, Inc., Sep 21, 1982