ClearanceBrief
Recall Z-0634-05 ยท CIZ Radioimmunoassay, Androstenedione

Diasorin Inc.

DiaSorin GammaCoat Androstenedione - [125I] RIA Kit Catalog No. CA-1725, DiaSorin Stillwater, Minnesota 55082-0285, U.S.A. Lot # 05144. for the quantitative determination of androstenedione levels in serum.

DateMar 18, 2005 (initiated Jul 15, 2004)
StatusTerminated
ReasonStandards used in the GammaCoat Androstenedione Radioimmunoassay Kit appeared yellowish in color and were gelatinous upon reconstitution.
Root causeOther
ActionCustomers were sent a customer notification letter identifying the product and the problem and requesting a response identifying the number of replacement kits needed.
Quantity163 Kits (lot 05144) 95 Kits (06114B)
DistributionMN, MA, KY, Saudi Arabia, Jordan, Sweden, Germany, Spain, Belgium, Italy, Canada, Australia, Portugal, Hungary, Switzerland & Israel
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K854650DSL ACTIVE ANDROSTENEDIONE RIA (DSL #3800)
Diagnostic Systems Laboratories, Inc., Dec 13, 1985