Recall Z-0676-05 ยท DHN Antinuclear Antibody, Indirect Immunofluorescent, Antigen, Control

Diasorin Inc.

PolyTiter software version 2.8 for use in the Polymedco PolyTiter Immunofluorescent Titration System, DiaSorin part # PT100 which is programmed for DiaSorin ANAFLUOR Indirect fluroescent Antibody Test For the detection and titration of circulating antinuclear antibodies. REF: 1624, 1662, 1660, 6624, 1604

DateApr 6, 2005 (initiated Jun 30, 2004)
StatusTerminated
ReasonThe program for a specific immunofluorescent titration system (Polymedco PolyTiter immunofluorescent Titration System) was found faulty in that the calibrator 2 value was incorrectly programmed as 1:120 instead of 1:160.
Root causeOther
ActionAll customers were notified of the problem via phone and letter. Verbal instructions were given of how to change the Calibrator 2 titer to the correct one. A letter instructs customer to return software disc to DiaSorin, remove version 2.8 from the computer and install version 2.8.1.
Quantity14
DistributionAL, AZ, CA, CO, FL, KY, MO, NC, NJ, NV, OH, TX, & VA