Recall Z-0703-05 ยท JTC Device, Microtiter Diluting/Dispensing
Diasorin Inc.
Version 7 Software for wash program on Bio Tek Automated Microplate Washer Model ELx50 which performs the washing steps for various programmed mictotiter plate assays.
| Date | Apr 13, 2005 (initiated May 20, 2004) |
| Status | Terminated |
| Reason | When software was updated for the Washer program of the Bio Tek Automated Microplate Washer Model ELx50, the software did not automatically clear old versions of the wash program when new ones were added. |
| Root cause | Other |
| Action | Customers were notified by phone and informed of the problem and recommended users to use only the newer wash program which is identified is a 'W' in front of the program. |
| Quantity | 9 |
| Distribution | CA, FL, NH, PA, SD, TX, WA |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K884125 | BIO-TEK AUTOMATED MICROPLATE WASHER MODEL EL403 Bio-Tek Instruments, Inc., Dec 2, 1988 |