Recall Z-0702-05 ยท MLM Enzyme Immunoassay, Tracrolimus

Diasorin Inc.

Version 12 software program written for the Bio-Tek ELx800 automated microtiter plate reader with use of PRO-Trac Tacrolimus ELISA kit

DateApr 13, 2005 (initiated Apr 12, 2004)
StatusTerminated
ReasonThe software program written for use in the automated microtiter plate reader (Bio-Tek ELx800) for use with the Pro-Trac II Tacrolimus ELISA kit was found to be faulty. Assay software program did not include the Calibrator Zero OD specification (STDO>=1.500) as specified in the product insert.
Root causeOther
ActionCustomers were sent notification letters and effectiveness check form. recommended users manually veify results using product insert specifications.
DistributionCA, Washington DC, FL, KY & Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.