Recall Z-0675-05 ยท JZO System, Test, Thyroid Autoantibody

Diasorin Inc.

Data reduction program software written specifically for Bio-Tek ELx800 (DiaSorin part # 15713) automated microtiter plate reader with use of Diastat Anti-Thyroglobulin Assay (manufacturer part #7700).

DateApr 6, 2005 (initiated Apr 12, 2004)
StatusTerminated
ReasonThe data reduction program written for a specific automated microtiter plate reader (Bio-Tek ELx800) was found to be faulty. This program is used to calculate the cut off values for the Diastat Anti-Thyroglobulin Kit assay. The negative Control Upper Limit should be less than 0.95, rather than less than 1.0.
Root causeOther
ActionCustomers were sent a notification letter directing customers to check historical data for negative control results reading greater than 0.95 but less than 1.0. These results are invalid and patient results should not have been reported. Customers are instructed to verify negative control manually until new software version is installed. An effectiveness check form is requested to be returned.
Quantity6
DistributionCA, FL, MA, NY & TX
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K952183DIASTAT ANTI-THYROGLOBULIN KIT
Shield Diagnostics, Ltd., Jun 30, 1995