Recall Z-0674-05 ยท MSV System,Test,Antibodies,B2 - Glycoprotein I (B2 - Gpi)

Diasorin Inc.

Data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800) reader) for use with Diastat Anti-Beta2 Glycoprotein ELISA assay kit.

DateApr 6, 2005 (initiated Oct 26, 2004)
StatusTerminated
ReasonData reduction Program written for specific sutomated microtiter plate insturment (Bio-Tek ELx800 reader) is faulty in that the positive result limit is to be greater than 15 U/ml, rather than greater than or equal to 15 U/ml.
Root causeOther
ActionCustomers were sent notification letters and effectiveness check form. recommended users manually veify positve results (RSLT), using product insert specifications.
Quantity7
DistributionCT, MI, MT, UT & CANADA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982879SHIELD DIASTAT ANTI-B2 GLYCOPROTEIN 1 IGG (ELISA) ASSAY
Shield Diagnostics, Ltd., Oct 15, 1998