Recall Z-0673-05 ยท DBM Antimitochondrial Antibody, Indirect Immunofluorescent, Antigen, Control

Diasorin Inc.

Data reduction program written specifically for Bio-Tek ELx800 automated microtiter plate reader with use of Diastat Anti-Mitochondrial Antibody Kit

DateApr 6, 2005 (initiated Apr 12, 2004)
StatusTerminated
ReasonThe data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800 reader) was found to be faulty. This program is used to calculate the cut off values for Diastat Anti-Mitochondrial Antibody Kit. Program should include a borderline specification to indicate repeat testing is needed.
Root causeOther
ActionA notification letter and effectiveness check form was sent to consignees. This letter recommended that users manually verify results using the specifications outlined in the package insert until a new software version is installed.
Quantity8
DistributionCA, CT, KS, MA & WA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K921183DIASTAT ANTI-MITOCHONDRIAL ANTIBODY KIT
Shield Diagnostics, Ltd., Jun 19, 1992