Recall Z-0495-05 ยท MKJ Automated External Defibrillators (Non-Wearable)

Medtronic Emergency Response Systems, Inc.

LIFEPAK 12 defibrillator/ monitors biphasic

DateFeb 3, 2005 (initiated Jan 15, 2005)
StatusTerminated
ReasonThe LIFEPAK 12 biphasic defibrillator/ monitors may deliver inappropriate energy levels and fail to escalate energy levels when configured to do so.
Root causeSoftware design
ActionThe firm sent 4000 User Notification letters via FED-EX on 1/15/2005 and approximately 3000 on 1/17/2005. The User Notification letters informed consignees about the two complaints and described the firm''s findings from their investigation into the problem. The letters also requested the users check the default energy settings; report back to the firm any discrepancies found; and reset the discrepant settings which differed from the user programmed settings.
Quantity28,835 devices distributed in the U.S.; 15,613 devices distributed internationally
Distribution28,835 devices were shipped to 6874 domestic consignees.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040775LIFEPAK 12
Medtronic Physio-Control Corp., Apr 23, 2004
K991910LIFEPAK 12 BIPHASIC
Physio-Control Corp., Sep 3, 1999