Recall Z-0527-05 ยท EKB Explorer, Operative

Hu-Friedy Mfg Co, Inc.

Dental Explorer #23 S/E, Mfr. Part Code/Product Code 100-3620, distributed under the Henry Schein label. Classification Name: 21CFR 872.4565; 510(k) exempt, Class I device, Device Listing # B057701. The product is labeled in part, HENRY SCHEIN 100-3620 Explorer-Single End #23 SE. +H658100362003. Made in China. For professional use only. Instruments must be cleaned and sterilized prior to use. Di

DateFeb 4, 2005 (initiated Jan 7, 2005)
StatusTerminated
ReasonPotential for tip breakage at the terminal shank due to improper heat-treatment at the foreign plant. Several reports of tip breakage were received by the own label distributor, and forwarded to the importer of record & specification setter of the dental explorers.
Root causeOther
ActionHu-Friedy Mfg. Co., Inc. notified the Own Label Distributor of the recall on 1/7/05 by phone, and in writing on 1/10/05. Henry Schein, Inc. is having formal "URGENT: DEVICE RECALL" letters printed up which will be mailed to over 8100 of Schein's direct customers by First Class Mail on/about 1/21/05 to initiate the sub-recall of the dental explorers. Henry Schein will follow-up to non-responders by phone and/or fax.
Quantity42,517 total units distributed (all date codes)
DistributionDental practitioners, hospitals, dental clinics and distributors located nationwide, in Puerto Rico and internationally.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.