Recall Z-0533-05 ยท DPS Electrocardiograph

Smiths Medical PM, Inc.

Advisor Vital Signs Monitor model 9200, catalog number 92E654325. Item Description 9200 II 3LD/RP/O2/IP/TP/BATT ME. The basic monitoring package includes ECG (3 lead / 5 lead), impedance respiration, non-invasive blood pressure, pulse oximetry, two invasive blood pressures, and tow temperature channels. The device permits patient monitoring with adjustable alarm limits as well as visible and a

DateFeb 10, 2005 (initiated Jan 5, 2005)
StatusTerminated
ReasonAn issue with the pace detect function was discovered on a 9200 II monitor. Could result in a falsely high or low heart rate reading.
Root causeOther
ActionOn 12/22/04 distribution of new units was placed on hold and customers were notified to recommend that they do not use the pace detect function. On 12/27/03 an email was sent to customers of the solution. On 12/29/04 a Technical Worksheet describing how to replace a resistor at R8 was emailed to customers. Between 12/29/04 and 01/03/05 kits containing instructions and new resistors were sent to customers to begin reworking of the devices.
Quantity1
DistributionInternationally only.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011177BCI ADVISOR VITAL SIGNS MONITOR (9200)
Bci, Inc., May 23, 2001
K030407ADVISOR, MODEL 9200
Bci, Inc., Aug 19, 2003
K982279VITAL SIGNS MONITOR; ADVISOR MODEL # BCI 9200
Bci Intl., Inc., Sep 25, 1998