Recall Z-0455-05 ยท LSL Dna-Reagents, Neisseria

Becton Dickinson & Co.

BD ProbeTec ET Urine Processing Kit, 4x25, catalog 440454, packaged in a carton containing 4 storage bags identified as BD ProbeTec ET Urine Processing Pouches, 25 pouches/bag. Carton and bags labeled in part ***Becton, Dickinson and Company, 7 Loveton Circle, Sparks, MD 21152***

DateFeb 3, 2005 (initiated Jan 10, 2005)
StatusTerminated
ReasonAn in-vitro diagnostic kit reagent may cause false negative clinical results in patients tested for sexually transmitted infections.
Root causeOther
ActionBD notified consignees by phone on 01/10/05 and advised that users transporting patient urine samples may experience increased indeterminate results or false negative results when testing the samples on the BD ProbeTec ET CT/GC Amplified DNA Assay. Subsequest to the phone notification, kit users identified as utilizing the Urine Processing Pouches were further notified by a letter. The letter describes 4 criteria contributing to the risk associated with false negative and indeterminate results. The letter advised users to discontinue use of the kits and discard for replacement. BD further ad
Quantity2175 kits
DistributionProduct was distributed to direct accounts in the US and to affiliated distributors internationally.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.