Recall Z-0440-05 ยท KNW Instrument, Biopsy

Ethicon Endo-Surgery

Mammotome EX Holster

DateJan 27, 2005 (initiated Dec 10, 2004)
StatusTerminated
ReasonA capacitor of the wrong value was used. The capacitor used was 0.11 microfarad instead of 0.22 microfarad called for in the product specifications.
Root causeOther
ActionThe recalling firm notified hospitals and distributors of the product by letter, dated 12/10/04.
Quantity120 devices
DistributionThe product was distributed to consignees located in MI, TN, IL, NC, GA, IL, CA, FL, IN, WI, TX, AR, KS, IA, OK, NY, PA, MS, KY, ME, MO, MA, VA, NJ, LA, MN, SD, UT, CO, AL, and NH. There are no foreign consignees.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033700MAMMOTOME EX HAND HELD SYSTEM
Ethicon Endo-Surgery, Inc., Feb 20, 2004