Recall Z-0341-05 ยท MKJ Automated External Defibrillators (Non-Wearable)

Philips Medical Systems

Philips Heartstart MRx Monitor/Defibrillator Models: M3535A (Hospital Use) , M3536A (EMS Use)

DateDec 21, 2004 (initiated Dec 15, 2004)
StatusTerminated
ReasonIn Software versions below A.02.00, with the sync function enabled and the device switched from manual to AED, there may be an unexpected delay in delivery of energy
Root causeOther
ActionPhilips notified customers on 12/13/04 by letter. Users are provided with action required to Mitigate Risk' until the correction is made by a Systems representative.
Quantity2520 units
DistributionNationwide Foreign:Canada, Austria, Australlia, Belgium, China, Denmark, Finland, France, Germany, Israel, Italy,Kuwait, Latvia, Lebannon, Libya, Netherland,Norway, Portugal, Qatar, Russia, Singapore,South Africa, Spain, Switzerland, Sweden, Taiwan, Thailand,Tunisia, UK, United Arab Emirates, Yemen
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031187HEARTSTART MRX, MODEL M3535A
Philips Medical Systems, Aug 27, 2003