ClearanceBrief
Recall Z-0353-05 ยท GBK

AGA Medical Corporation

AMPLATZER Delivery System 45-degree curve, order no. 9-DEL-8F-45/60. Product is labeled as Sterile EO. Amplatzer Occluder Device and Delivery System is comprised of delivery sheath, delivery cable, dilator, loading device and pin vise. The Delivery and Exchange Systems are placed onto polyboard backing cards. The backing card is then placed into a smaller of 2 Tyvek pouches. The pouch is then

DateDec 24, 2004 (initiated Nov 23, 2004)
StatusTerminated
ReasonTubing utilized in the manufacture of the AMPLATZER Delivery and Exchange System sheaths and dilators was in some cases manufactured by our supplier outside of process parameters; causing a residue to be left on the inside of the tubing.
Root causeOther
ActionRecall notices were sent via e-mail 11/22/04 to physicians, hospitals and distributors who received affected product. Hard copies of the Recall Notices were sent to all customers (international and domestic) via Federal Express 11/22-24/04. A form that is attached to the Recall Notice asks for customers to fill out quantities of affected product and to return the affected product to AGA Medical.
Quantity1750
DistributionProduct has been distributed throughout the US and Internationally, including Canada, Europe and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.