Recall Z-0538-05 ยท DAP Fibrinogen And Fibrin Split Products, Antigen, Antiserum, Control

Dade Behring , Inc.

Advanced D-Dimer. Fibrinogen and fibrin split products Catalog numbers OQWM11 and OQEM13.

DateFeb 15, 2005 (initiated Dec 3, 2004)
StatusTerminated
Reasondecrease in product stability after reconstitution and reduced stability on board certain instrument systems
Root causeOther
ActionThe recalling firm issued a recall letter to their customers informing them of the problem and the need to use internal quality controls with the analyzers. The letter instructed customer to discard lot # 544545 and reduce the stability for the other lot numbers involved.
Quantity25381 units
DistributionThe product was shipped nationwide to medical facilities.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041438ADVANCED D-DIMER ASSAY
Dade Behring, Inc., Aug 11, 2004