Recall Z-0339-05 ยท GKZ Counter, Differential Cell

Beckman Coulter Inc

ACT diff 2 Analyzer (Automated differential cell counter)

DateDec 21, 2004 (initiated Dec 3, 2004)
StatusTerminated
ReasonErroneous results issue when MONOJECT blood collection tubes are used during analysis after tubes are uncapped and recapped then analysed in the closed vial mode. Results are lower than if cap had not been removed.
Root causeOther
ActionA letter issued to users informing them of the incompatability of the Coulter ACT diff 2 Analyzer with the use of Monoject blood collection tubes made by Tyco Healthcare Group/Kendall Healthcare. Customers choosing to use these tubes are instructed to uncap the tubes and analyze the samples,as open vial samples.
Quantity2709
DistributionNationwide and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K990352COULTER AC T DIFF 2 ANALYZER
Coulter Corp., Apr 1, 1999