Recall Z-0387-05 ยท DCK C-Reactive Protein, Antigen, Antiserum, And Control

Ortho-Clinical Diagnostics

Vitros CRP (C-Reactive Protein) Slides. Catalog #192 6740 contains 250 slides; Catalog #809 7990 contains 90 slides. Firm on label: Ortho-Clinical Diagnostics, a Johnson & Johnson Company, Rochester, NY 14626.

DateJan 8, 2005 (initiated Nov 22, 2004)
StatusTerminated
ReasonResults may be negatively biased due to a non-linear response at the upper end of the Reportable (Dynamic) Range.
Root causeOther
ActionLetters dated 11/22/2004 issued to customers. End users instructed to manually modify the Reportable and/or Analyzer Ranges; dilute all samples with an initial CRP result greater than 9.0 mg/dL. Distributors instructed to destroy product in-house and to notify end-users to manually modify the Reportable and/or Analyzer Ranges.
QuantityCat. 1926740 - 20,903 boxes; Cat 8097990 - 34,877 boxes
DistributionNationwide and worldwide distribution. 70% of product was shipped to foreign consignees.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030626VITROS CHEMISTRY PRODUCTS CRP SLIDE AND CALIBRATOR KIT 7
Ortho-Clinical Diagnostics, Inc., Mar 18, 2003