Recall Z-0367-05 ยท FII System, Dialysate Delivery, Sealed

Baxter Healthcare Renal Div

System 1000 family of Hemodialysis Instruments, including the System 1000, Arena, 1550, BM11, BM25 and Meridian; Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A.

DateDec 24, 2004 (initiated Dec 14, 2004)
StatusTerminated
ReasonAir bubbles have been observed in the tubing sets past the air detector of the instrument with no alarms. Air bubbles in the circuit can cause an air embolism.
Root causeOther
ActionAn Urgent Device Correction letter dated 12/14/04 was sent along with a reply form and addenda to operator''s manuals to all Baxter customers who have purchased hemodialysis hardware. The letter informed the accounts of reports of air bubbles observed in the tubing sets past the air detector without an alarm alerting the operator of potential air in the tubing. A review of the reports revealed misuse, including failure to follow generally observed hemodialysis practices. Baxter updated their Hemodialysis Operator''s Manuals and Training Guides to clarify instructions and add caution and warni
DistributionThis action covers Baxter''s complete line of hemodialysis hardware, which was distributed nationwide, including Puerto Rico, and internationally to Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Denmark, Ecuador, Finland, France, Republic of Georgia, Greece, Guatema
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030099BAXTER ARENA HEMODIALYSIS DELIVERY SYSTEM, MODEL 1571278000
Baxter Healthcare Corp, Jun 17, 2003
K872364HEMODIALYSIS (SPS), MODEL 550, CODE 5M1309
Travenol Laboratories, S.A., Oct 13, 1987
K894838HEMODIALYSIS SINGLE PATIENT SYSTEM (SPS) 550 VSB
Baxter Healthcare Corp, Sep 7, 1989
K970446DRAKE WILLOCK SYSTEM & ALTRA TOUCH 1000 DIALYSIS DELIVERY SYSTEM
Althin Medical AB an Affiliate of Baxter Intl, Jun 2, 1997