Recall Z-0385-05 ยท FPA Set, Administration, Intravascular
Baxter Healthcare Corp.
Baxter labeled 6060 Solution Sets with AutoClamp; For use with model 6060 multi-therapy pumps only; a prescription, sterile, non-pyrogenic fluid pathway; Manufactured by an affiliate of Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., the following sets were made in Costa Rica: a) Baxter Solution Set, 76'' (1.9 m), Volume 3.7 mL, Cassette with AutoClamp Device, Male Luer Lock Adapter, p
| Date | Jan 8, 2005 (initiated Nov 19, 2004) |
| Status | Terminated |
| Reason | A missing platen assembly in the cassette of the solution sets results in the infusion pump being unable to occlude the tubing of the set. This can lead to free-flow of the solution and a possible overdose of medication to the patient. |
| Root cause | Other |
| Action | Urgent Product Recall letters dated 11/19/04 were sent to the direct accounts on the same date via UPS ground. The accounts were informed that free-flow situations were associated with the listed product codes and lot numbers due to a missing platen assembly in the cassette of the solution sets, which could result in an overdose of medication to the patient. The accounts were requested to stop use of the affected lot numbers and product codes and return the sets to Baxter c/o NNC Group, 2670 Executive Dr., Indianapolis, IN 46241, following the enclosed instructions. Any questions concerning |
| Quantity | 3,063,359 sets |
| Distribution | Nationwide, including Puerto Rico, and internationally to Australia, Belgium, Canada, Hong Kong, Israel, Japan, Singapore, Sweden and Switzerland |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K002679 | 6060 MULTI-THERAPY PUMP, MODEL 2M9832; 6060 EPIDURAL PUMP, MODEL 2L9351 Baxter Healthcare Corp, Sep 18, 2000 |