Recall Z-0299-05 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Baxter Healthcare Renal Div

Baxter Meridian Hemodialysis Instrument, product codes 5M5576 and 5M5576R; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085 U.S.A.

DateDec 8, 2004 (initiated Dec 2, 2004)
StatusTerminated
ReasonThe instrument power cord may become disconnected from the instrument and pose a fire or electrical shock hazard.
Root causeOther
ActionBaxter sent Urgent Device Correction letters dated 12/2/04 to all Meridian customers on the same date. The letters informed the accounts of the potential for disconnection of the power cord and teh potential for arcing, fire or electrical shock. The accounts were requested to examine their instruments to see if a new power cord clamp has been installed on the unit, using the enclosed instructions and complete the enclosed reply sheet indicating the number of units already corrected, and the number of units still requiring the power cord clamp. The accounts were requested to fax the complete
Quantity3,004 units
DistributionNationwide and internationally to Mexico, China and Korea.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K992894MERIDIAN HEMODIALYSIS MACHINE, MODEL 5M5576
Baxter Healthcare Corp, Nov 24, 1999