ClearanceBrief
Recall Z-0297-05 ยท NPQ Test, Factor V Leiden Mutations, Genomic Dna Pcr

Roche Diagnostics Corp.

MagNA Pure LC Instrument; catalog numbers 2236931 and 03670325001.

DateDec 8, 2004 (initiated Nov 5, 2004)
StatusTerminated
ReasonTest results may be adversely affected by software shutting down the cooling blocks in systems with software version 3.09.
Root causeOther
ActionConsignees were notified via letter dated 11/5/04, which provides the user with a workaround.
Quantity65
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.