Recall Z-0272-05 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips Medical Systems (Cleveland) Inc

Intinion 1.5 T MRI Scanner.

DateDec 2, 2004 (initiated Apr 17, 2003)
StatusTerminated
ReasonThere is a potential for the patient to be exposed to cryogenic fluids (liquid air) dripping from the vent during a quench due to the location of the vent.
Root causeOther
ActionA Mandatory Service Letter, dated April 17, 2003, was sent to service technicians instructing the technicans to perform a mandatory safety modification.
Quantity93 units.
DistributionThroughout the United States and to Denmark, England, China, Italy, Mexico, Israel, and Puerto Rico.