ClearanceBrief
Recall Z-0454-05 ยท DQY Catheter, Percutaneous

Boston Scientific Scimed

RunWay 6F Cardiovascular Guide Catheter.

DateJan 28, 2005 (initiated Oct 29, 2004)
StatusTerminated
ReasonThe Runway guide catheter do not meet internal product specifications. Specifically, the affected catheters have exposed or protruding braid wires in the inner lumen of the catheter tip. This condition could have the potential to damage a POBA or Stent Delivery System balloon
Root causeOther
ActionOn 10/29/04 the Boston Scientific Sales force was notified via voice mail of the recall. Sales personnel were told to contact their accounts/customers by the end of the business day on 11/02/04. Accounts/customers were to receive written notice of the recall via Fed Ex on Monday 11/01/04. Sales Personnel are to ensure that accounts/customers got notice of the recall, understand instructions and are taking appropriate steps. Catheters are to be returned to Boston Scientific Corporation, Quincy, MA.
Quantity76,984
DistributionThroughout United States and into 15 other countries.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033441BOSTON SCIENTIFIC 6F RUNWAY GUIDE CATHETER
Boston Scientific Corp, Nov 21, 2003