Recall Z-0371-05 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter Inc

UniCel Dxl 800 Access Immunoassay Systems (Reaction Vessels).

DateDec 28, 2004 (initiated Oct 29, 2004)
StatusTerminated
ReasonBeckman Coulter has confirmed reports of rare occurrences of UniCel Dxl 800 Access Immunoassay Systems Reaction Vessels that are deformed or flattened in their packaging bags for lot 04250160. Deformed or flattened RVs could present a rish of causing jams on the UniCel Exl 800 Access Immunoassay Systems that may require service intervention and could compromise sample analysis.
Root causeOther
ActionAll customers that were shipped lot 042500160 of the reaction vessels were contacted via telephone (or via on-site visit - Canada only) October 29 - November 1, 2004. A Product Correction Action Letter dated November 2,2 004, sent via U.S. mail to all customers, once contacted by phone with fax back forms.
Quantity56 boxes
DistributionFL, HI, IL, MS, MI, NC, NY, OH, OK, TX, UT, VA, WI, and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023764UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM, MODEL 973100
Beckman Coulter, Inc., Jan 28, 2003