Recall Z-0556-05 ยท NPJ Prosthesis, Knee Patellofemorotibial, Partial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Stryker Howmedica Osteonics Corp.

The Triathlon Posteriorly Stabilized Femoral Component.

DateFeb 24, 2005 (initiated Oct 27, 2004)
StatusTerminated
ReasonThe fatigue testing of the Triathlon PS Femoral Component does not consistently meet the requirements of the finite element analysis predicted load as delineated in the 510(k) submittal.
Root causeOther
ActionNotification letters and Product Accountability Froms have been sent via Fedex on 10/27/2004 with return receipt.
Quantity386 UNITS
DistributionThe product is distributed to its branches/agencies and to hospitals nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.