Recall Z-0262-5 ยท CHI Lipase-Esterase, Enzymatic, Photometric, Lipase

Beckman Coulter Inc

SYNCHRON Clinical Systems Lipase (LIPA) Reagent; Classification name: 862.1465

DateNov 30, 2004 (initiated Oct 25, 2004)
StatusTerminated
ReasonBeckman Coulter has confirmed that occasionally a cuvette may be skipped during the SYNCHRON Lipase Wash (LIWA) procedure. If this occurs, there is the potential for carryover into the SYNCHRON Lipase (LIPA) assay, resulting ina substantial positive bias affecting Lipase (LIPA) results.
Root causeOther
ActionThe week of 10/25/2004, a product corrective action (PCA) letter was mailed vis U.S. mail to all customers shipped LIPA reagent for the 24 months preceding the discovery of the issue. Customers instructed to repeat any sample that produces a lipase value of > 170 U/L that cannot be confirmed by patient history.
Quantity6,698
DistributionUS and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K970905SYNCHRON CX SYSTEMS LIPASE (LIPA) REAGENT
Beckman Instruments, Inc., Jun 4, 1997