Recall Z-0958-05 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems (Cleveland) Inc

AcQSim X-Ray Computed Tomography Scanners. These scanners were sold under the brand names: Picker, Marconi or Philips.

DateJul 6, 2005 (initiated Mar 23, 2004)
StatusTerminated
ReasonSoftware anomaly in the DICOM Modality Worklist. When restoring a patient's images from the optical disk, images from the wrong patient may be displayed.
Root causeOther
ActionThe firm mailed a ''Product Safety Notification'', dated 3/23/2004, to all users who had DICOM Modality Worklist enabled with a software key. The customers were notified of the problem and were provided instructions as to how to prevent the anomaly from occurring. The notification also informed the users that updated software which corrects the anomaly will be available in the near future and will be installed by the firm''s Field Service Engineers.
Quantity44
DistributionNationwide and Austria, Italy, Netherlands, Spain, United Kingdom, and Sweden. The units were installed in medical facilities located in the following states within the U.S.: KS, PA, FL, MN, IA, CA, GA, LA, MS, OH, NE, MO, IN, TN, TX, MI, NM, MA, AL and NH;
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001088ULTRASIM CT SYSTEM
Philips Medical Systems (Cleveland), Inc., Jun 22, 2000