Recall Z-0208-05 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corp

Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required); Model 801763.

DateNov 18, 2004 (initiated Aug 12, 2003)
StatusTerminated
ReasonThe central control monitor screen may go blank during use because of a bad wiring connection in the wiring harness or a low voltage setting on the potentiometer.
Root causeOther
ActionService representatives were sent a service bulletin dated 8/12/03 instructing them to correct units in the field bearing serial numbers between 20 and 109.
DistributionUnited States, Belgium, Canada, Dubai, Egypt, Germany, Hong Kong, India, Japan, Korea, Singapore and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022947ADVANCED PERFUSION SYSTEM 1 (100/120 VOLT),ADVANCED PERFUSION SYSTEM 1 (220/240 VOLT, ROLLER PUMPS (6, 4 INCH
Terumo Cardiovascular Systems Corp., Mar 20, 2003