Recall Z-0269-05 ยท IXR Table, Radiographic, Tilting

General Electric Med Systems LLC

Patient tables of CT systems with brand name: CT/e; CT/e Dual; CT/e Plus, CT/e Dual Plus; CT/e Lite; ProSpeed AI; ProSpeed AII; ProSpeed FI; ProSpeed FII; ProSpeed EII; CT HiSpeed Series.

DateFeb 24, 2005 (initiated Oct 21, 2004)
StatusTerminated
ReasonThe patient tables of the affected CT systems are supported by a hydraulic cylinder and a cylinder shaft (i.e. the upper shaft) which connects the tabletop to the table frame. GE received a report of a HiSpeed LX/I system table falling abruptly due to the filure of the upper shaft. No injuries were reported.
Root causeOther
ActionInadequate materials used to manufacture the upper shaft of patient tables that can result in uncontrolled table motion.
Quantity1150