Recall Z-0188-05 ยท FKX System, Peritoneal, Automatic Delivery

Baxter Healthcare Renal Div

HomeChoice and Yume Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A.; catalog numbers 5C4471, 5C4471R, T5C4441, T5C4441R

DateNov 11, 2004 (initiated Nov 1, 2004)
StatusTerminated
ReasonThe on/off power switch is not grounded as a separate component in the device, and can loosen from its housing, which may result in an electrical shock to the user.
Root causeOther
ActionBaxter sent Urgent Device Correction letters to the Peritoneal Dialysis Home Care Coordinators on 11/1/04 and to the Home Dialysis Patients on 11/2/04. The consignees were informed of the potential for a shock if the power switch loosens in the housing, and that the switch would be replaced with a grounded power switch on a next service call basis. If the patient experiences any difficulty turing the device on or off, or if they notice that the power switch is loose, they were instructed to immediately unplug the device and contact Baxter Global Technical Service (BGTS) at 1-800-553-6898, op
Quantity12,200 units
DistributionNationwide, including Puerto Rico, and internationally to Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Ecuador, Guatemala, Honduras, Hong Kong, India, Japan, Korea, Malaysia, Nicaragua, New Zealand, Peru, Philippines, Saudi Arabia, Singapore, South Africa, Taiwan, Turkey, Uruguay an