Recall Z-0196-05 ยท IZI System, X-Ray, Angiographic

Siemens Medical Solutions USA, Inc

POLYSTAR T.O.P., Angiographic X-Ray System, Model Numbers 1148902 and 4784505.

DateNov 16, 2004 (initiated Sep 23, 2004)
StatusTerminated
ReasonOrbital gears may exhibit excess wear
Root causeOther
ActionThe recalling firm issued Update Instructions, AX077/04/S, to Siemen''s Customer Service Representative (CSR) to visit affected sites and check the orbital gears for excessive wear. If the CSR finds gears exhibiting excessive wear the orbital gears will be replaced. If no wear is observed the gears will be lubricated as a preventive measure.
Quantity82 units
DistributionThe product was shipped to medical facilities nationwide and government accounts in KS and MI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K913120POLYSTAR ANGIOGRAPHIC X-RAY SYSTEM
Siemens Medical Solutions USA, Inc., Aug 16, 1991